Lead Lined Wood Doors manufactured to national woodwork standards, constructed with the highest quality timber core, Rail, Stile and face veneers available.All lead lined doors, must provide a lead shielding equivalent of equal or greater value as specified by the project physicist of record or other "qualified experts" radiation shielding calculations report recommendations and requirements. Customer must verify required lead thickness values and equivalents prior to ordering.
Doors must be internally shielded with lead covering full height and width of door
Radiation safety and general medical construction guidelines require as a "good practices" rule that the gap between the bottom of an x-ray door and the finished floor should always be kept as small as possible.
Door lockset hardware should be lead lined if lead value requirement is more than 1/16" thick, or if specified.
Door must have interlock switch so that x-ray machine cannot be activated when door is open or ajar.
Application
High-strength Wooden door with 2 mm lead sheet inside to prevent the leak of X-ray.
Laminate or Veneer can be pasted as per requirement.